Instruments > USB-Powered Potable Potentiostats
pocketSTAT2.LC

Ultra-low current sensitivity, in an ultra-small design

The pocketSTAT2.LC is a portable potentiostat/galvanostat including an integrated Low Current Module with sub-femtoampere current resolution.

±10pA real current range
Integrated EIS 1MHz
Field or lab use
Request Info
Lowest native current range of ±10pA with resolution down to 0.3 fA, the pocketSTAT2.LC can measure the finest of details.
Unlock non-destructive insights into complex reaction kinetics and surface properties with EIS - integrated in the pocketSTAT2.
This USB-powered compact and leightweight device can go anywhere you can take your laptop.

Applications

Specifications

Key Specs

Maximum current
±30mA
Compliance voltage
±10V
Maximum applied potential
±10V
EIS frequency range
10μHz to 1MHz
Electrode Connections
3: WE, CE, RE (+GND)
Maximum sampling rate
300,000 samples/s
IR compensation
Yes

Potentiostat

Maximum applied potential
±10V
Applied potential accuracy
0.2% or 2mV
Current ranges
±10pA to ±10mA
Measured current resolution
0.003% of current range, minimum 0.3fA
Measured current accuracy
20pA + 0.025% FSR

Galvanostat

Applied current ranges
±10pA to ±10mA
Applied current accuracy
0.2%
Applied current resolution
0.008% of applied current range
Potential ranges
±1mV, ±4mV,±10mV,±40mV, ±100mV,±400mV, ±1V,±4V, ±10V
Measured potential accuracy
0.2% or 2mV
Measured potential resolution
0.0008% of potential range, minimum 7nV

Impedance Analyser

EIS Frequency range
10μHz to 1MHz
EIS Amplitude
0.15mV to 2.0V or 0.03% to 100% of current range

IviumSoft

Free and included powerful electrochemical software

pocketSTAT2.LC

Size
w x d x h = 22.3 x 6.7 x 1.9cm
Weight
350g

Requirements

Power requirements
via USB
Interfacing
USB
PC requirements
Windows 8 or higher

Includes

Special low current cell cable USB connection cable Electrode clips Test cell Carry case Full IviumSoft electrochemical software

Get in touch

We're here to help with:

  • Instrument pricing and quotations
  • Selecting the right instrument
  • Technical questions
  • Hardware and software support
  • Help and advice

Electron Transfer Recorded! Thanks for your message, we'll get back to you in 1-2 days
Volatge Drop Detected! Something went wrong with the form. Please try again, or send us an email info@redcatscientific.com.au

425 Smith Street
Fitzroy
Vic 3065
Australia

info@redcatscientific.com.au
(+61) 0483694373

Redcat Scientific 2026